Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Parts

812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557

812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557

  • 812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557
  • 812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557
  • 812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557
  • 812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557
  • 812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557
812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557
Product Details:
Brand Name: Mindray
Model Number: 812A-30-08557 812A-20-08558 V.B
Payment & Shipping Terms:
Minimum Order Quantity: 1pcs
Packaging Details: Carton
Payment Terms: T/T(wire transfer), Western Union, Paypal
Contact Now
Detailed Product Description
Delivery Time: 3-5 Working Days Product Name: 812A-30-08557 812A-20-08558 V.B Mindray PM7000 Patient Monitor ECG Board
Brand: Mindray Part No: 812A-30-08557 812A-20-08558 V.B
Use For: Mindray PM7000 Patient Monitor MOQ: One Piece
Warranty: 90 Days Condition: Work In Good Condition
Delivery Method: DHL,UPS ,Fedex Payment Terms: In Full
High Light:

812A-20-08558 Patient Monitor ECG Board

,

812A-30-08557 Patient Monitor ECG Board

,

ECG Board Patient Monitor Parts

812A-30-08557 812A-20-08558 V.B Mindray PM7000 Patient Monitor ECG Board

 

 

A brief introduction to the maintenance and troubleshooting of monitors

 

 

Methods of Troubleshooting and Troubleshooting for ECG Monitor

 

The ECG monitor plays an important role in the whole monitoring process. Since the monitor works continuously for almost 24 hours, its failure rate is also high. The common faults and troubleshooting methods are introduced as follows.

 

 

1. No display when booting

 

Fault phenomenon: When the instrument is turned on, there is no display on the screen and the indicator light is not on; when the external power supply is connected, the battery voltage is low and the alarm is turned off, and then the machine automatically shuts down; useless.

 

 

Inspection Method:

① When the instrument is not connected to AC, check whether the 12V voltage is low. This fault alarm indicates that the output voltage detection part of the power board detects a low voltage, which may be caused by a fault in the detection part of the power board or a fault in the output of the power board, or it may be caused by a fault in the back-end load circuit.

 

 

②When the battery is installed, this phenomenon indicates that the monitor is working on the battery power supply and the battery power is basically used up, and the AC input is not working normally. The possible reason is: the 220V power socket itself has no power, or the fuse is blown.

 

 

③When no external battery is connected, it may be judged that the rechargeable battery is broken, or the power board/charging control board is faulty and the battery cannot be charged.

 

 

Troubleshooting method: connect all the connecting parts reliably, connect the AC to charge the instrument.

 

 

2. ECG has no waveform

 

Symptoms of failure: connect the lead wire but there is no ECG waveform, and the display shows "electrode off" or "no signal received".

 

Checking method: first check the lead mode, if it is five-lead mode but only three-lead connection is used, there must be no waveform.

 

Secondly, on the premise of confirming the placement position of the ECG electrode and the quality of the ECG electrode, exchange the ECG cable with that of other machines to confirm whether the ECG cable is faulty, whether the cable is aging, or the pin is broken. .

 

Again, if the failure of the ECG cable is ruled out, the possible cause is that the "ECG signal line" on the parameter socket board is not in good contact, or the ECG board, the main control board connection line of the ECG board, and the main control board are faulty.

 

 

 

Product Name

Mindray PM7000 Patient Monitor ECG Board 12A-30-08557

 

Brand Mindray
Part No 12A-30-08557
Use for Mindray PM7000 patient monitor
MOQ One piece
Warranty 90 days

 

 

 

Elimination method:

 

(1) Check all the external parts of the ECG leads [the three/five extension lines in contact with the human body should be connected to the corresponding three/five contact pins on the ECG plug, if the resistance is infinite, it means that the lead wires are broken. , the lead wire should be replaced].

 

812A-20-08558 V.B Patient Monitor ECG Board 812A-30-08557 0

(2) If the waveform channel of the ECG display shows "no signal reception", it means that there is a problem with the communication between the ECG measurement module and the host, and the prompt still exists after turning off and on again, and you need to contact the supplier.

 

 

Thank you for your interest ,if you have any interest in 812A-30-08557 812A-20-08558 V.B Mindray PM7000 Patient Monitor ECG Board ,feel free to contact us .Thank you .

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Kiara

Tel: 8619854815217

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.