Send Message
Tel:
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group

China World Technology Medical Equipment Service Group

Win-win together!

Home ProductsPatient Monitor Accessories

Nellcor DEC-8 Pulse Oximetry SpO2 Extension Cable For Welch Allyn Vital Signs Monitor 300 Series

Nellcor DEC-8 Pulse Oximetry SpO2 Extension Cable For Welch Allyn Vital Signs Monitor 300 Series

  • Nellcor DEC-8 Pulse Oximetry SpO2 Extension Cable For Welch Allyn Vital Signs Monitor 300 Series
  • Nellcor DEC-8 Pulse Oximetry SpO2 Extension Cable For Welch Allyn Vital Signs Monitor 300 Series
  • Nellcor DEC-8 Pulse Oximetry SpO2 Extension Cable For Welch Allyn Vital Signs Monitor 300 Series
  • Nellcor DEC-8 Pulse Oximetry SpO2 Extension Cable For Welch Allyn Vital Signs Monitor 300 Series
Nellcor DEC-8 Pulse Oximetry SpO2 Extension Cable For Welch Allyn Vital Signs Monitor 300 Series
Product Details:
Brand Name: Nellcor
Model Number: DEC-8
Payment & Shipping Terms:
Minimum Order Quantity: 1
Delivery Time: 3-5 working days
Payment Terms: T/T, Western Union,PayPal
Contact Now
Detailed Product Description
Product Name: Nellcor SpO 2 Extension Cable Type: DEC-8
Application: Medical Equipment,ICU,hospital Device Warranty: 90days
Color: Blue Stock Situation: Available Now
Service: Online Technical Support,Passion In Serving Our Customer With The Best Service MEANS OF TRANSPORT: Express Delivery
Preponderance: Advanced Equipment,Original Minimum Purchase Commitment: 1pcs
Scenario: ICU, Emergency, ICU Room And Operating Room,etc Product Specialty: Easy To Use,Economic,Functional
Power Supply Source: Electricity Power Consumption: 110V—240v
Material: Plastic,etc
High Light:

Pulse Oximetry SpO2 Extension Cable

,

SpO2 Extension Cable

 

Nellcor DEC-8 pulse oximetry SpO 2 extension cable for Welch Allyn Vital Signs Monitor 300 Series

 

 

► Brief Introduction

Verify that the sensor cable connector has slots on both sides. If the cable connector has a slot on only one side, then the sensor is not compatible with the monitor.

The SpO 2 connectors are notched and flanged to ensure proper fit. If the connectors do not join easily, stop and verify the following:

  • You have the correct SpO 2 sensor.
  • The cable connector is aligned correctly.

 

Test connections

Test connections Connector/cable Connect to Wiring Connect to
SpO 2 (Nellcor) NellcorD-connector(DEC-8 or other OXI-MAX equivalent) SpO 2 Input Connector All wires connected together and terminated appropriately for your hi-pot tester High Voltage

 

 

Testing the monitor only (Nellcor SpO 2 )

Use this procedure to test only the monitor SpO 2 function.

1. Power the monitor off.

2. Connect the Nellcor SRC-MAX SpO 2 functional tester to the SpO 2 input connector through a Nellcor DEC-8 extension cable.

3. Power the monitor on.

4. Verify the following on the SRC-MAX:

  • All of the monitor LEDs flash: left panel, center panel, right panel.
  • SRC-MAX initializes to a default condition where the four test parameter LEDs are lit closest to their selector buttons
  • The default pulse rate is 60 bpm and the default SpO 2 is 75%

5. Give the monitor up to 30 seconds to stabilize, and verify a displayed pulse rate of 60 ± 3 bpm and a displayed SpO 2 of 75 ± 3%.

6. Set the SRC-MAX pulse rate to 200 bpm.

7. Give the monitor up to 30 seconds to stabilize, and verify a displayed pulse rate of 200 ± 3 bpm.

8. Switch the SRC-MAX SpO 2 saturation percentage to 90.

9. Give the monitor up to 30 seconds to stabilize, and verify a displayed SpO 2 saturation level of 90 ± 3%.

10. Disconnect the SRC-MAX.

 

 

 

Test the isolation between SpO2 and DC input:

  1. Connect the SpO 2 test cable between the SpO 2 connector and the high-voltage jack on the hi-pot tester.
  2. Push the test button to run the AC Withstand Voltage test.
  3. Verify that the monitor passed the test. (If the input current never exceeds 2.50 mA, the isolation is sufficient.)
  4. Remove the SpO 2 test cable.

Nellcor DEC-8 Pulse Oximetry SpO2 Extension Cable For Welch Allyn Vital Signs Monitor 300 Series 0

 

Company Profile

We mainly sell different kinds of medical equipments, accessories and repair parts including patient monitor, defibrillator machine, patient monitor printer, ECG machine, EKG machine, main board, batteries and so on. After many years of marketing and developing, we are quiet proficient in medical machines technology.

 

Our products are increasingly popular and enjoy good reputation at home and abroad. Relying on superior products quality and reasonable price, our sales have been growing steadily with each passing year. We are a reliable manufacturer providing the best products and services for global customers. Satisfying all the customers' demands is our permanent pursuit.

 

 

 

Looking forward to receiving your early inquiry about Nellcor DEC-8 pulse oximetry SpO 2 extension cable for Welch Allyn Vital Signs Monitor 300 Series. If you have other questions about our products or services, please feel free to contact us. We are always ready to serve you.

Contact Details
China World Technology Medical Equipment Service Group

Contact Person: Rita

Tel: 8618279091952

Send your inquiry directly to us
China World Technology Medical Equipment Service Group
TianHe District,GuangZhou,GuangDong
Tel:86-198-5481-5217
Mobile Site Privacy Policy | China Good Quality Used Patient Monitor Supplier. © 2021 - 2024 patients-monitor.com. All Rights Reserved.